Recall Z-0133-2021 ยท FQO Table, Operating-Room, Ac-Powered

Maquet Cardiovascular Us Sales, Llc

Yuno II Mobile Operating Table - Product Usage: support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Part 143302F0 (US) Part 143302B0 (OUS)

DateSep 24, 2020
StatusTerminated
ReasonPlastic socket of the IEC 60320-1 socket pulled out of the metal cage and electronic parts exposed; where the socket was pulled out and main cable is still connected to the main socket (power source), there is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components
Root causeUnder Investigation by firm
ActionGetinge issued Urgent Medical Device Recall on September 24, 2020, via FedEx 2 day delivery to US consignees. The letter states reason for recall, health risk and action to take: Per the IFU, section 9.1 (Yuno) and 8.1 (Yuno II), a visual and functional check is required to be performed by a trained person before operation. Follow the instructions in section 4.2.3.1of the Yuno and Yuno II IFU regarding connecting and disconnecting from the mains socket (power source). If the plastic socket of the cold-device plug is pulled out, do not try to put it back into the housing and if there are any ex
Quantity6 units US
DistributionWorldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, KENYA, KOREA, KUWAI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.