Recall Z-0398-2021 ยท GAM Suture, Absorbable, Synthetic, Polyglycolic Acid

Gordian Surgical

TroClose 1200 - Product Usage: intended for use in a variety of gynecologic, general and urologic endoscopic procedures, to create and maintain a port of entry and to facilitate the delivery of absorbable sutures and anchors through soft tissues of the body during endoscopic / laparoscopic surgery. The trocar may be used with or without visualization for primary and secondary insertions. It is com

DateSep 22, 2020
StatusTerminated
ReasonEndotoxin test results found above the acceptable levels in LAL test (above 20EU/device).
Root causeProcess control
ActionAll associated LOTs shall be accounted for and all devices in those lots. Since only two LOTs are affected and only two costumers/distributors are in possession of said LOTs no failed attempt to correspond with them is expected. If any of the distributors fails to respond, a Gordian costumer representative will personally go to location to follow through on the process indicated.
Quantity300 total devices
DistributionUS Nationwide distribution including in the states of Texas and Michigan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160564TroClose1200
Gordian Surgical, Ltd., Nov 17, 2016
K171494TroClose1200, TroClose1200 Cannula
Gordian Surgical, Ltd., Jun 15, 2017