Recall Z-0128-2021 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

System, Nuclear Magnetic Resonance Imaging - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.

DateSep 8, 2020
StatusTerminated
ReasonPotential electrical arc flash within the Power Gradient & RF (PGR) Cabinet.
Root causeProcess change control
ActionOn 09/08/20 the firm sent a Customer Letter to its consignees with the following instructions: You can continue to use the system. In the event the PGR cabinet needs to be serviced, power down the PGR at the Main Disconnect Panel (MDP) then follow MDP Lockout/Tagout from the current released service methods and documentation. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction
Quantity30 devices total
DistributionWorldwide distribution - US Nationwide distribution including in the states of FL, KY, NC, NJ, NM, NY, TX, UT and WA. The countries of Canada, Hungary, India, Korea, China, Russia, Taiwan, Puerto Rico and Ubekistan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091536OPTIMA MR450W
Ge Medical Systems, LLC, Jul 17, 2009
K163331Discovery MR750 3.0T; Discovery MR750w 3.0T;Discovery MR450 1.5T; Discovery MR450w 1.5T; SIGNA Architect and SIGNA Artis
Ge Medical Systems, LLC, Mar 17, 2017