Recall Z-0705-2021 ยท FPA Set, Administration, Intravascular
ICU Medical, Inc.
30" (76 cm) Appx 5.5 mL 20 Drop Admin Set w/Integrated Clave Drip Chamber, 1.2 Micron Filter, Spiros, Bag Hanger, REF: CH3286
| Date | Aug 31, 2020 |
| Status | Open, Classified |
| Reason | Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line. |
| Root cause | Process control |
| Action | On 08/31/2020, Medical Device Recall notices were mailed to customers informing them to take the following actions: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) ICU Medical has some lots of unaffected product available today and |
| Quantity | 325 |
| Distribution | Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countrie |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K941190 | CLAVE NEEDLE-LESS CONNECTOR OPEN EXTENSION SET Icu Medical, Inc., Jan 31, 1995 |