Recall Z-0056-2021 ยท HAW Neurological Stereotaxic Instrument

Brainlab AG

Ultrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x and 3.1.x - Product Usage: intended to be an intra-operative image guided localization system to enable minimally invasive surgery.

DateAug 31, 2020
StatusTerminated
ReasonBrainlab Ultrasound Navigation Software does not support the modification of the probe's image width.
Root causeSoftware design
ActionThe intraoperative Brainlab Ultrasound Navigation Software does not support the modification of the probes image width, an optional feature of the BK Medical Ultrasound System, resulting in an incorrect display of the depth of the ultrasound overlay by the navigation for non-linear ultrasound probes, with the deviation depending on the probe and the difference between the calibrated and used image width.
Quantity271 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of Illinois, Massachusetts, Texas, California, Idaho, Missouri, Oregon, Connecticut and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Curacao, Czech Republic, Egypt, Finland, France, Germany, Switzerla
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092467VECTORVISION CRANIAL, VECTORVISION ENT, KOLIBRI CRANIAL, KOLIBRI ENT, CRANIAL ESSENTIAL, CRANIAL UNLIMITED
Brainlab AG, May 6, 2010