ClearanceBrief
Recall Z-0284-2021 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips North America, LLC

Philips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A, M4744A - Product Usage: used in pairs to transmit a controlled electrical shock from an external defibrillator directly to the exposed heart muscle of a post-pubescent patient in order to intentionally stop start the heartbeat during cardiopulmonary surgery.

DateSep 14, 2020
StatusTerminated
ReasonThe periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has created an addendum to the IFU, that includes a test for the switched paddles only, which will detect this failure mode.
Root causeDevice Design
ActionUrgent Medical Device Correction notification letters dated 9/4/20 were sent to customers. Insert a copy of the attached Instructions for Use Addendum: Sterilizable Defibrillator Paddles: Insulation Resistance Check for Switched Internal Paddles to the Sterilizable Defibrillator Paddles IFU. Follow the Instructions for Use, Paddles Checks section. The Paddle Checks activities include: Mechanical Check, Visual Inspection, Functional Check, Continuity Check, and Insulation Resistance Check. Perform these activities to confirm the paddles are safe and ready for use. The Insulation Resistance T
Quantity8,942
DistributionGlobal Distribution. US Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031187HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Aug 27, 2003