ClearanceBrief
Recall Z-0293-2021 ยท OUG Medical Device Data System

Pentax of America Inc

9310HD Digital Video Capture Module with software version 3.4.0 or higher configured with 9263 endoPortal

DateOct 23, 2020 (initiated Sep 22, 2020)
StatusTerminated
ReasonThere is an intermittent software issue that could affect the systems, in which an exam video for one patient (Patient A) might be copied to another patient (Patient B). This does not occur at the time the user performs and initially reviews an exam; it is not evident until a follow-up review occurs.
Root causeSoftware design
ActionURGENT MEDICAL DEVICE CORRECTION notification letters were sent on 9/22/20. Customer Instructions: Enclosed with this letter is a Field Correction Response Form which identifies the serial numbers of the affected products your facility owns. PENTAX Medical will be taking a two step approach to contain and correct the problem. o First, we will contact you to schedule a time for our Field Service to temporarily disable file transfer to endoPortal. o During the time that file transfers to endoPortal are disabled, you can continue to create and save exam video files. When this issue is resolve
Quantity39 total
DistributionUS Nationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.