Recall Z-0125-2021 ยท FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days

Becton Dickinson & Company

BD Nexiva Closed IV Catheter System Dual Port (20 GA 1.75 IN) Catalog (Ref) No. 383538 BD Nexiva closed IV catheter systems are intended to be inserted into a patient s vascular system for short term use to sample blood, monitor blood pressure, or administer fluids.

DateSep 14, 2020
StatusTerminated
ReasonDull/blunt needles within the IV Catheter System may lead to missed insertion attempts and delay in diagnosis or treatment until a new needle is inserted
Root causeNonconforming Material/Component
ActionSiemens issued an "URGENT MEDICAL DEVICE RECALL" letter dated 9/14/20 to Distributors/Medical Facilities on 9/14/20 via FedX , email or faxed stating reason for recall, health risk and action to take: 1. Immediately review your inventory for the specific Catalog (Ref) and Lot number listed above. Destroy all product subject to the recall following your institutions process for destruction. 2. Share this recall notification with all users within or outside your facility network to which you may have further distributed the affected product to ensure they are also aware of this recall. 3. Compl
Quantity12,020 units
DistributionUS: Nationwide Distribution: AK, AR, AZ, CA, IL, KS, MO, NY, OH, UT, VA; and International to: Canada and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183399BD Nexiva Closed IV Catheter System
Becton Dickinson Infusion Therapy Systems, Inc., Mar 20, 2019