Recall Z-0702-2021 ยท FPA Set, Administration, Intravascular

ICU Medical, Inc.

30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger, REF: 20124-01; Spinning Spiros Closed Male Luer, Red Cap, REF: 20130-01; Spinning Spiros, Closed Male Luer, REF: CH2000S; Spinning Spiros Closed Male Luer, 10 units, REF: CH2000S-10; Spinning Spiros Closed Male Luer, Red Cap, REF: CH2000S-C; Spiros, Closed Male Luer w/Red Cap, 10 Units, RES: CH2000SC-10

DateAug 31, 2020
StatusOpen, Classified
ReasonPotential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Root causeProcess control
ActionOn 08/31/2020, Medical Device Recall notices were mailed to customers informing them to take the following actions: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) ICU Medical has some lots of unaffected product available today and
Quantity1856578
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countrie
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082806SPINNING SPIROS, MODEL CH-2000
Icu Medical, Inc., Apr 6, 2009