ClearanceBrief
Recall Z-0081-2021 ยท MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer

Biomet, Inc.

Biomet Regenerex Primary Taper Cap Item Number 141269

DateSep 4, 2020
StatusTerminated
ReasonProduct not properly being aligned with the adequate gamma sterilization group and result in the product not being properly sterilized
Root causeProcess change control
ActionZimmer Biomet notified on 9/2/20 Distributors with direct sales and/or fielded inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory, as well as hospitals that have consumed products in the past 12 months will be notified via FedEx. Surgeons that have consumed products with in the past 12 months notified via FedEx. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quaran
QuantityUS: 62 units; OUS: 176 units
DistributionNationwide Foreign: Canada, FINLAND, NETHERLANDS, UNITED KINGDOM
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080361REGENEREX TIBIAL COMPONENTS
Biomet, Inc., Apr 21, 2008