ClearanceBrief
Recall Z-0093-2021 · PHX Shoulder Prosthesis, Reverse Configuration

Biomet, Inc.

Biomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate with Taper Adapter-Shoulder Prosthesis item: 010000589 - Product Usage: intended to serve as the foundation for the reverse shoulder prosthesis head (glenosphere).

DateSep 2, 2020
StatusTerminated
ReasonProduct potentially being packaged without a taper adapter or baseplate. The package should contain one (1) baseplate and one (1) taper adapter, product may be packaged with two (2) baseplates and no (0) adapters or with no (0) baseplates and two (2) adapters
Root causePackaging process control
ActionZimmer Biomet issued Urgent Medical Device Recall Letter dated 9/02/20 to Distributors via email and Hospital Risk Mangers via courier stating reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have unconsumed affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to Corporat
Quantity3 units
DistributionUS Nationwide distribution including in the states of CO, MN, OH, WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120121COMPREHENSIVE REVERSE SHOULDER-MINI BASEPLATE
Biomet, Inc., Feb 16, 2012
K193373Comprehensive® Reverse Shoulder
Biomet, Inc., Apr 15, 2020