ClearanceBrief
Recall Z-0290-2021 ยท MIH System, Endovascular Graft, Aortic Aneurysm Treatment

W L Gore & Associates, Inc.

GORE TAG CONFORMABLE Thoracic Stent Graft with ACTIVE CONTROL System

DateSep 9, 2020
StatusOpen, Classified
ReasonDue to secondary deployment fiber being attached to secondary deployment handle, completion of deployment via the deployment line access hatch was not possible
Root causeUnknown/Undetermined by firm
ActionOn 09/23/2020, the firm sent an "URGENT Medical Device Safety Correction" via First Class Postal (USPS) informing them that all lots of its Thoracic Stent Graft Systems may encounter inability to complete secondary deployment and that the Recalling Firm will be updating it Instruction for Use to include a new Warning: Secondary deployment events have occurred where the device does not open to full diameter after Secondary Deployment Handle actuation, and the deployment line is not accessible via the Deployment Line Access Hatch. If this event occurs, Gore recommends using the GORE Tri-Lobe Bal
Quantity13860 units
DistributionU.S. - Nationwide O.U.S. - Australia, Austria, Belgium, Croatia, Cypress, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Lebanon, Luxembourg, Monaco, Netherlands, Norway, New Zealand, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.