ClearanceBrief
Recall Z-0177-2021 ยท ERL Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece

Bien-Air Surgery Sa

Bien-Air Surgery REF 1600517-001 SN ******* MULTIFUNCTION PEDAL OSSEODUO - Product Usage: Product Usage: intended to convert electrical energy to mechanically drive attachment, microsaws, shaver blades and burs, burs and perforator tools for cutting and shaping bone including spine and cranium.

DateSep 11, 2020
StatusOpen, Classified
ReasonThe magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.
Root causeProcess control
ActionOn 09/14/2020, the firm send out an "URGENT: Medical Device Correction" notification informing customer of the potential for the magnet located inside the foot pedal displaces or comes off during a surgery and therefore may compromise the device's safety. The use of the incriminated devices may entail a risk of unattended start or unstoppable motor when acting on the foot pedal. The Recalling Firm is informing customers of a curative and a corrective action to prevent the magnet from coming off: -Curative Action: 1) Inspect your inventory for the product numbers/serial numbers above. 2) Quar
QuantityTotal US = 1120 foot pedals
DistributionUS Nationwide distribution including in the states of CA, MI and MN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083720OSSEODUO, SHAVER HANDPIECE S80 AND S120, MICROMOTOR 80K
Bien-Air Surgery SA, Mar 20, 2009