Recall Z-3023-2020 ยท OWB Interventional Fluoroscopic X-Ray System

Philips North America, LLC

ProxiDiagnost N90

DateNov 20, 2019
StatusTerminated
ReasonUnexpected increase in peak tube potential (kV) will lead to an increased patient radiation dose.
Root causeRadiation Control for Health and Safety Act
ActionPhilips sent an Urgent Field Safety Notice dated August 5, 2020 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Philips plans to correct the defect at no cost to customers by implementing a field correction involving the installation of a software update. A Philips Field Service Engineers will schedule an appointment with customers to install the software update.
Quantity71
DistributionUnited States
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173433ProxiDiagnost N90
Philips Medical Systems Dmc GmbH, Feb 5, 2018