Recall Z-0115-2021 ยท NCW System, Test, Her-2/Neu, Monitoring
Siemens Healthcare Diagnostics, Inc
Atellica IM HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10995591, UDI Codes: (01)00630414599151(10)98034139(17)20210130 and (01)00630414599151(10)32478139(17)20210130 Atellica IM HER-2/neu (H2n) calibrator, 2 Pack, Catalog No. SMN 10995592, UDI Code (01)00630414599168(10)22695A49(17)20201211 ADVIA Centaur HER-2/neu (H2n) assay, 50 Test Kit, Catalog No. SMN 10308994, UDI codes : (01)00630
| Date | Aug 25, 2020 |
| Status | Terminated |
| Reason | Siemens confirmed that the Upper Limit of Normal (ULN) as claimed in the Instructions for Use (IFU) was no longer achieved. This can potentially result in false positive or false negative results. |
| Root cause | Process control |
| Action | All affected customers were sent a letter on 08/25/2020 stating the following: Siemens Healthcare Diagnostics Inc. (Siemens Healthineers) notified customers on August 25th, 2020. Letter states reason for recall, health risk and action to take. Actions to be taken by their customers include: - Reviewing the letter with their Medical Director. - Customers may continue use of the product lots referenced in Table 1 until they receive replacement product in their laboratory. Refer to the Table 3 for product biasinformation. - If you are currently using the products listed in Table1, review your |
| Quantity | Atellica IM H2n Assay =135 kits; Atellica IM H2n Calibrator = 46 kits; ADVIA Centaur Assay = 1,823 kits; ADVIA Centaur |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |