ClearanceBrief
Recall Z-0057-2021 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Zimmer Biomet, Inc.

Dual Mobility Vivacit-E Bearing, Model Number 110031009

DateAug 14, 2020
StatusTerminated
ReasonThe outer package is labeled as a Size C 38 mm, however, the implant inside the package is a Size E 42 mm.
Root causeUnder Investigation by firm
ActionUrgent Medical Device Recall notification letters dated 8/14/20 were sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not ha
Quantity31
DistributionProducts were distributed to the following US states: FL, ID, IL, IN, KY, MA, MD, MI, MN, NC, NJ, OH, OR, PA, SC, TX, UT, VA, WI, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190656Dual Mobility - Longevity and Vivacit-E Polyethylene Hip Bearings
Zimmer, Inc., Dec 6, 2019