Recall Z-0085-2021 ยท KPA Slide Stainer, Automated

Roche Diagnostics Operations, Inc.

BenchMark ULTRA Advanced Staining System (IVD)- Material Number: 05342716001

DateAug 31, 2020
StatusOpen, Classified
ReasonPotential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the electrical cord connector and socket
Root causeDevice Design
ActionRoche issued Urgent Medical Device Correction letter on 8/31/20 via UPS Ground Receipt stating: reason for recall, health risk and action to take: actions below. " Inform any operators that use the BenchMark ULTRA and DISCOVERY ULTRA instruments of the potential hazards associated with this issue and provide a copy of this notification as appropriate. " Inspect the instrument daily for evidence of leaks (i.e., liquid outside of the waste reservoir, liquid on the floor, liquid on the skin/outer panel of the instrument). " If you experience a fluid leak, stop using the system and turn off the po
Quantity2623 (US); 3735 (ex-US)
DistributionNationwide Foreign: Canada, Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.