Recall Z-0613-2021 ยท HLM Instrument, Measuring, Lens, Ac-Powered

Carl Zeiss Meditec AG

Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038 - Product Usage: is a device for measuring the refractive power of contact lenses and spectacle lenses and displaying measurement values for sphere, cylinder and prism.

DateAug 6, 2020
StatusTerminated
ReasonDue to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.
Root causeNonconforming Material/Component
ActionOn 08/18/2020, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via FedEx overnight informing customers that based on post market surveillance activities, the Recalling Firm found that the communication board on their digital lensmeter and autorefractor were not performing as expected and needs to be exchanged as a precautionary measure. In rare cases the RS232 communication port can fail on the communication board. Both products use the same technology. In the worst-case situation, the safety function of the communication board is insufficient when a connected third-party
Quantity603 units
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TX, UT, VA, WA and WY. The countries of Argentina, Austria, Australia, Belgium, Bolivia, C
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.