Recall Z-3006-2020 ยท CGI Radioimmunoassay, Estriol

Beckman Coulter Inc.

BECKMAN COULTER ACCESS Immunoassay Systems - Access Unconjugated Estriol, REF 33570.

DateAug 26, 2020
StatusTerminated
ReasonDue to alkaline phosphatase (ALP) associated interference causing falsely elevated uE3 (Unconjugated Estriol assay) results.
Root causeUnder Investigation by firm
ActionOn 08/31/2020, the firm sent an "URGENT MEDICAL DEVICE RECALL" notification to customers via email and US Post informing them of the potential for an interference with bovine alkaline phosphatase (ALP) in a subset of patient samples tested with the Access Unconjugated Estriol (uE3) assay. Customers are instructed to: - Consider the patients total clinical presentation when interpreting results. - Any retrospective review of patient results is left to the discretion of the Medical Director. - Contact Beckman Coulter Technical Support if you suspect ALP associated interference and do not report
Quantity21455 assay systems
DistributionWorldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, NC, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, WA, WI and Puerto Rico. The countries of Australia, Belgium, Bosnia and Herzegovina, Brazil, Canad
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.