Recall Z-0087-2021 ยท DXF Catheter, Septostomy

Medtronic Vascular

Rashkind Balloon Septostomy Catheter, 6F, Recess Balloon, Closed end, 50 cm, Pediatric, Single Lumen Product number: 008764 GTIN: 00613994760289 The Rashkind balloon septostomy catheters are designed for creating an atrial septal defect or enlarging an existing atrial septal defect as a treatment option for patients with a Cyanotic Congenital heart defect. Cyanotic congenital heart defects are ra

DateOct 16, 2020 (initiated Aug 25, 2020)
StatusTerminated
ReasonSubsequent failure of catheters in the field and failed quality testing
Root causeDevice Design
ActionAn immediate field corrective action was initiated by Medtronic field representatives. Customers were notified verbally to quarantine and return affected product on 25-Aug-2020. The following information was verbally provided to affected customers by Medtronic representative: "A supplier for Rashkind Catheter product has notified Medtronic that it has shipped product to Medtronic that may contain an issue. The supplier conducts a routine test in production. In two separate lots, the manufacturing specification was not met by all samples, but the lots were inadvertently shipped to Medtronic. W
Quantity2356 units
DistributionWorldwide Distribution: USA (nationwide) states of: MA, PA, TN, IN, MI, WA, CA, CO, AZ, NC; and OUS countries of: Algeria, Armenia, Australia, Austria, Barbados, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Finland, France, French Polynesia, Germany, Greece, Hungary, India, Ireland, Israel, I
FDA recordOpen on fda.gov