Recall Z-2970-2020 ยท CGA Glucose Oxidase, Glucose

Siemens Healthcare Diagnostics Inc

Siemens epoc BGEM Test Card-In vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care SMN#/Model #: 10736382

DateAug 28, 2020
StatusTerminated
ReasonSporadically inconsistent discrepant (low bias) glucose results on card lot 01-20095-10. Potential exists for a delay in diagnosis of hyperglycemia or unnecessary treatment for hypoglycemia.
Root causeProcess control
ActionSiemens issued Urgent Medical Device Correction letter on August 28, 2020 via Federal Express stating the reason for recall communication letter informs customers of the issue and, health risk and action take: Please dispose any inventory of test card lot 01-20095-10 currently in your possession in accordance with local and state disposal requirements. You may request free of charge replacement product from your local Siemens or distributor office. Please review your inventory of these products and assess your laboratorys replacement needs. Please use epoc BGEM Test Card from another unaffec
Quantity463 boxes (50 test cards/box )
DistributionNationwide Foreign: Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200107epoc Blood Analysis System
Epocal, Inc., Jun 5, 2020