Recall Z-0119-2021 ยท FBN Choledochoscope And Accessories, Flexible/Rigid

Olympus Corporation of the Americas

Olympus CHF-CB30S Choledoscope Intended Use: The device was designed to be used with an Olympus TV camera system, light source, TV monitor, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis and treatment within the biliary tract (common bile duct, cystic duct, and hepatic duct).

DateAug 31, 2020
StatusTerminated
ReasonThe CHF-CB30S Device did not have a 510(k) clearance. Additionally, the device has been associated with breakage of the insertion section and protrusion of metal parts, as well as breakage and displacement of the rubber on the bending section during surgical procedures.
Root causeDevice Design
ActionAll affected customers were sent a "URGENT: MEDICAL DEVICE REMOVAL ACTION" letter on 08/31/2020 stating the following: Olympus Corporation notified direct customer accounts on 08/31/2020 via recall notification letter. Letter states reason for recall, health risk and action to take. Actions to be taken by customers including the following: 1. Inspect your inventory and identify any CHF-CB30S devices. 2. Cease any further use of any CHF-CB30S device you have, remove it from your inventory and quarantine it until it is shipped back to us. 3. Call your Olympus customer service representative at
Quantity193 devices distributed in the U.S. 548 devices distributed globally.
DistributionUS Nationwide Distribution in states of: AK, AZ, CA, FL, GA, IL, LA, MA, ME, MI, MN, MS, NC, NM, NJ, NY, OR, SC, TN, WA, and WI; and International to: Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.