Recall Z-3025-2020 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

Software versions syngo.CT VB20 or VB20_SP1 in the following systems: SOMATOM Force (Model #10742326) SOMATOM Definition As (Model #8098027) SOMATOM Definition Edge (Model #10590000) SOMATOM Definition Flash (Model #10430603) SOMATOM Drive (Model #10431700) SOMATOM Confidence (Model #10590100) SOMATOM Edge Plus (Model #10267000)

DateSep 28, 2020 (initiated Jun 22, 2020)
StatusTerminated
ReasonSporadic problems with the current software may result in scanning workflow interruptions and unexpected user notifications causing delay in diagnosis or patient rescan
Root causeSoftware design
ActionSiemens Medical Solutions USA, Inc. has initiated a Customer Information Letter (CIL) to affected customers via CT004/20/S. The letter alerts customers of the technical issues and under what circumstances they may occur and health risk These issues will be resolved with software update version syngo CT VB20_SP2 (Service Pack 2) via UI CT003/20/S. The software update is being made available to all affected customers free of charge. If your system is connected to Siemens Remote Connection, this update will be performed remotely. The remote software update process will require approximately 2 ho
Quantity767 distributed in U.S.
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103127SOMATOM DEFINITION AS
Siemens Medical Solutions USA, Inc., Mar 4, 2011
K121072SOMATOM DEFINITION FLASH
Siemens Medical Systems, Inc., May 8, 2012
K133589SOMATOM FORCE
Siemens Medical Solutions USA, Inc., Apr 17, 2014
K142955SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition AS/AS+, SOMATOM Definition AS Open,
Siemens Medical Solutions USA, Inc., Nov 24, 2015
K161196SOMATOM Drive
Siemens Medical Solutions USA, Inc., Aug 24, 2016
K173630SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM Definition AS Open, SOMATOM Dei
Siemens Medical Solutions USA, Inc., Mar 30, 2018