Recall Z-0054-2021 ยท OCC Respiratory Virus Panel Nucleic Acid Assay System

Luminex Corporation

VERIGENE Processor SP Model No. 10-0000-07. An in vitro diagnostic device intended for processing and identifying nucleic acids

DateDec 10, 2018
StatusOpen, Classified
ReasonNo results/incorrect results due to failure of the hybridization heater.
Root causeComponent design/selection
ActionOn or about 11/19/2018 there was one complaint from a customer regarding a thermal control failure. In response to the complaint the Field Service Engineer replaced hybridization heater and ran all tests as passed. Subsequently, the customer reported the initial service was ineffective. The VERIGENE SP Processor was then removed from the customer site and returned for analysis. No Customer Notification was communicated for this removal.
Quantity1 system
DistributionUS distribution in MO. No OUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143653Verigene Respiratory Pathogens Flex Nucleic Acid Test (RP Flex)
Nanosphere, Inc., Sep 4, 2015