ClearanceBrief
Recall Z-2997-2020 ยท KPS System, Tomography, Computed, Emission

GE Healthcare, LLC

Discovery NM670 Pro Model # H3100XB

DateAug 14, 2020
StatusTerminated
ReasonRotor bearing screws were found loose on detector.
Root causeUse error
ActionGE Healthcare issued an Urgent Medical device correction notification on August 14, 2020. It states the following: 1. GE Healthcare will inspect and, if required, correct all affected products at no cost to you. 2. A GE Healthcare representative will contact you to arrange for the correction 3. Please complete the acknowledgement form and return it to GE Healthcare promptly upon receipt and no later than 30 days from receipt to nm.fmi40890.responses@ge.com If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local S
Quantity9 units
DistributionUSA : AR CT FL GA IL KS KY LA MA MD MI MN MO NE NJ NV NY OH OK RI SC TN TX VA VT WA WI OUS: ARGENTINA AUSTRIA BELGIUM BRAZIL Canada CHINA FRANCE GERMANY GREECE INDIA ISRAEL ITALY JAPAN Korea (Republic of) NETHERLANDS NORWAY PANAMA POLAND PORTUGAL Russia SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND T
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093514DISCOVERY MODEL NM/CT 670
Ge Medical Systems F.I. Haifa, Dec 10, 2009