ClearanceBrief
Recall Z-2945-2020 ยท HRS Plate, Fixation, Bone

Smith & Nephew, Inc.

EVOS Screws - Product Usage: is a part of the EVOS MINI plating system, indicated for fracture fixation, arthrodesis, reconstruction, replantation or reduction of small bones and small bone fragments.

DateAug 24, 2020
StatusTerminated
ReasonMultiple or incorrect expiration dates were provided on the affected designation labels.
Root causeError in labeling
ActionOn August 24th the firm sent via email a notification to affected consignees. Consignees are asked to inspect inventory, locate and quarantine affected devices. Sales Reps and or distributors are instructed to notify customers of the field action and ensure required actions are complete. Affected product will be returned to Smith & Nephew. All consignees are asked to complete a Response Form, even if they do not have product to return.
Quantity71 units
DistributionInternational Distribution to the countries of Denmark, Sweden, France and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140814EVOS MINI-FRAGMENT PLATING SYSTEM
Smith & Nephew, Inc., May 7, 2014