ClearanceBrief
Recall Z-3043-2020 ยท DIO Enzyme Immunoassay, Cocaine And Cocaine Metabolites

Ameditech Inc

QuickTox Drug Screen DipCard(AMT Item/ AMT Description): QT70/ "QuickTox Drug Screen Dipcard 11 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+OXY100+MDMA500", QT88/ "QuickTox Drug Screen Dipcard 12 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+OXY100+ MDMA500+BUP10", QT11 / "QuickTox Drug Screen Dipcard 5 Drug Detection: COC300+OPI

DateAug 18, 2020
StatusTerminated
ReasonVarious immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Root causeProcess control
ActionA Field Corrective Action Notice letter was sent to affected consignees. The Customer notifications commenced on 19-Aug-2020. The consignees are requested to return all non-expired product in their possession to the recalling firm. Method of Notification, 1st contact: Written field corrective action notification letters will be sent via certified mail with delivery tracking to each consignee along with a pre-paid return label to return any product remaining in stock. 2nd & 3rd contact: Telephone, follow up email, or written mail communication will be used as necessary.
Quantity170,425 units
DistributionUS - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.