Recall Z-3015-2020 ยท IKY Mattress, Flotation Therapy, Non-Powered

Stryker Medical Division of Stryker Corporation

Stryker IsoAir 2.0 support surface-provides pressure redistribution and is intended to assist in management of the microclimate of the patient skin, Model Number: 2941-001-100

DateAug 12, 2020
StatusTerminated
ReasonConnector between support surface and pump may be out of tolerance resulting in difficulty in removing the CPR connector causing a delay in resuscitation efforts for patients experiencing cardiac arrest and needing CPR
Root causeNonconforming Material/Component
ActionStryker issued Medical Device Correction Notice dated 8/12/20, initially in person or via email by a Stryker Sales Representative with the customer notification letter. Letter states reason for recall, health risk, action to take: 1. Locate the units listed on the attached business reply form and remove them from use until they can be corrected. 2. Return the enclosed business reply form to confirm receipt of this notification by fax 269 488 8691 or email productfieldaction@stryker.com. 3. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the correction
Quantity110 units
DistributionUS Nationwide distribution including in the states of CA, FL, NY, OH.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.