ClearanceBrief
Recall Z-2975-2020 ยท CEM Electrode, Ion Specific, Potassium

Inpeco S.A.

Aptio Automation systems with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Clinical Chemistry Test Systems - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system

DateAug 7, 2020
StatusTerminated
ReasonFirmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
Root causeSoftware design
ActionInpeco issued Customer Letter dated August 7, 2020 forwarded to the customers either by Inpeco (for the laboratories where Inpeco is directly the distributor) or by the Inpeco distributors. Letter states reason for recall, health risk and action to take: customers impacted by the Field Action have to complete and return the Recall Letter Receipt Confirmation and Implementation Check form attached to the Customer letter within 15 days to Inpeco or the distributor. The customers impacted by the issue will be contacted by Inpeco or the distributors to arrange a visit to upgrade the impacted modu
Quantity10 US and 22 OUS
DistributionWorldwide distribution - US Nationwide distribution including in the states of AK, FL, IA, LA, MA, MS, MT, NC, OH, TX, VA and the countries of AUSTRALIA, AUSTRIA, BELGIUM, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, NETHERLANDS. NORWAY SAUDI ARABIA SOUTH KOREA SWEDEN UNITED ARAB EMIRATES
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121012FLEXLAB 3.6, ACCELERATOR A3600
Inpeco S.P.A., Aug 31, 2012