Recall Z-2934-2020 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA, Inc.

Symbia S-Series and Symbia T-Series - Product Usage: Intended use of the device SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.

DateAug 14, 2020
StatusTerminated
ReasonThis letter is to inform you of the possibility that your Symbia S or Symbia T system(s) may have unintended slow motion of the detectors in specific situations which could pose a potential risk.
Root causeDevice Design
ActionOn 8/14/20 the firm notified their customers via letter of the field action, with the following instructions: How can you help to avoid the potential risk of this issue? You may continue to use your system ensuring to follow all instructions in the Operator Manual. Do not leave the patient unattended at any time while the patient is positioned on the patient bed. Siemens Medical Solutions USA, Inc. 2501 N. Barrington Road Hoffman Estates, IL 60192 USA In the event of an Emergency-Stop, shutdown or power loss the patient should be manually retracted from the gantry using the patient bed palle
Quantity412 units
DistributionUS Nationwide distribution including in the states of Alabama, Alaska, Arizona, California, Colorado, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Missouri, Nebraska, New Jersey, New York, North Car
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200474Symbia 6.7
Siemens Medical Solutions USA, Inc., Mar 20, 2020