Recall Z-0541-2021 ยท NQM Immunoassay For Detection Of Amniotic Fluid Protein(S).

Clinical Innovations, LLC

Clinical Innovations ROM Plus Collection Kit IFU: ROM Plus Rupture of Membrane Test ROM-5025, ROM-6025

DateAug 13, 2020
StatusTerminated
ReasonPossible false negative or false positive results. The incorrect sample collection swabs (non-scored) were assembled into assembled into the in vitro diagnostic kits. If the user does not break the swab stick correctly, or if scissors or other tools are used to cut the swab handle false negative or false positive results can occur.
Root causeProcess change control
ActionOn August 17, 2020 through August 19, 2020, the firm sent an "URGENT SAFETY NOTICE" to customers via email informing them that the incorrect swabs (non-scored) were package with their in vitro diagnostic kits and that new instructions and training material will be provided. Customers are instructed to identify any affected products and to complete the response card that is attached to the Customer Notification Letter. Customers are also provided the option of returning the affected products to the Recalling Firm. A company representative will contact and arrange for the return of the produc
Quantity19450 units/kits
DistributionUS: CO, IL, IN, MO, OH, RI, UT, and VA. OUS: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110605ROM PLUS
Clinical Innovations, LLC, Nov 23, 2011