Recall Z-0108-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis zee or Artis Q systems (listed below) - Product Usage: for single and biplane diagnostic imaging and interventional procedures. System Name/Material Number: Artis Q BIPLANE 10848282 ARTIS Q CEILING 10848281 Artis Q floor 10848280 ARTIS Q ZEEGO 10848283 ARTIS Q.ZEN BIPLANE 10848355 ARTIS Q.zen ceiling 10848354 Artis Q.zen floor 10848353 Artis zee Biplane 10094141 Artis zee ceiling 10094137 A

DateAug 3, 2020
StatusTerminated
ReasonPotential exists for the collision supervision not to work properly following loss of the individual room configuration data settings causing danger of collision with fix mounted room equipment, walls or floor may occur. This might cause components of the system to fall or tilt which could result in crushing of patients, operators or staff, collision of system parts with patients, as well as result in delay or interruption of the clinical procedure
Root causeSoftware Manufacturing/Software Deployment
ActionSiemens iMedical Solutions USA Inc. issued A Customer Safety Advisory Notice dated 8/03/20 affected customers via AX047/20/S. Letter states reason for recall, health risk and action to take:Siemens strongly recommends paying close attention during system movements to avoid any collision and to establish appropriate emergency procedures until the corrective action has been performed. In case of collision, it is recommended to check the system for any damages before continuing with treatment. If any device part continues to move although that movement was not released, a potential malfunction
Quantity124 U.S
DistributionUS Natonwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181407Artis zee/zeego & Artis Q/Q.zen
Siemens Medical Solution USA, Inc., Aug 15, 2018