Recall Z-2990-2020 ยท NSU Instrumentation For Clinical Multiplex Test Systems

Hologic, Inc

Panther Fusion Tube Tray, Part: FAB-15004, REF: PRD-04000

DateAug 4, 2020
StatusTerminated
ReasonTube trays may leak which could invalidate patient results or assay worklists, causing delayed results for any assays with which they are used. A possibility exists of false negative result with flu A/B/RSV assay or paraflu assay.
Root causeUnder Investigation by firm
ActionOn 08/04/20, Medical Device Recall Notifications were mailed to Laboratory Managers, Site Administrators, and Operators. Customers were asked to do the following: 1. Discontinue use. 2. Immediately check inventory and segregate affected devices. 3. Complete the customer response form. Document the remaining inventory. Destroy the segregated inventory. Return the customer response form even if there is no affected inventory present. 4. There is no other specific actions to the assays. 5. Recalling firm will ship replacement product based on the customer response form information.
Quantity607
DistributionU.S. :ID, SC, VT, CA, IL, UT, NJ, MA, TX, AL, CO, FL, ME, NY, VA, WA, WI, OR, DC, GA, NC, HI, CT, OK, TN, O.U.S.: CA, SE, GB, DE, DO, IT, ES, NO, BE, NL, FR, DK, AT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.