Recall Z-2953-2020 ยท GFD Dermatome

Integra LifeSciences Corp.

13 ft Handpiece Cable, component of the Dermatome Sets Model S and Model S6 and also as a spare part - Product Usage: are intended for use in the cutting of skin grafts for use in plastic surgery and/or burn treatment. These instruments are intended only for professional use in hospitals and similar medical facilities, where the patient will be under the supervision of trained personnel.

DateSep 12, 2020 (initiated Aug 4, 2020)
StatusTerminated
Reasonlntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
Root causeMixed-up of materials/components
ActionAll impacted consignees will be notified by letter delivered by courier service, facsimile or email about the initial notification between August 4th and 6th, 2020. Customers are asked to remove the affected product from use and complete and return the response form to the manufacturer. After response forms are reviewed the customer will be asked to return the product and a replacement will be sent.
Quantity74 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of MN, TX, CA, UT, WV, FL, IN, GA, AZ, and the countries of Latin America ( CR, MX, TT, EC, PA ), Canada, Europe (FR, AT, BE, ES).
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.