ClearanceBrief
Recall Z-2966-2020 ยท JIF Enzymatic Method, Ammonia

RANDOX LABORATORIES, LTD.

AM1054 Randox Ammonia (NH3), Enzymatic UV Method For the quantitative in vitro determination of Ammonia in plasma

DateJul 30, 2020
StatusTerminated
ReasonRandox Ammonia reagents are being recalled from the field due to a positive bias of up to 140mol/l being observed on patient samples.
Root causeUnder Investigation by firm
ActionThe firm, Randox Laboratories Ltd., sent an "Urgent Medical Device Correction" letter dated 30th July 2020 to it customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: - Discontinue use of, quarantine and discard any stock of AM1015, batch 517541 and AM1054, batch 512103 immediately. - Review results generated with the affected batches in line with the clinical profile of the patient. - Discuss the contents of this notice with your Medical Director. -Review your reagent inventory of these products and assess your labor
Quantity143 kits in total
DistributionUS Distribution to states of: VA, UT, WV and country of : Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030873RANDOX AMMONIA
Randox Laboratories, Ltd., May 12, 2003