Recall Z-2968-2020 ยท JJX Single (Specified) Analyte Controls (Assayed And Unassayed)

Tosoh Bioscience Inc

Tosoh Bioscience AIA-PACK BHCG Calibration Verification Test Set-In Vitro Diagnostic for BHCG-pregnancy Part number: 020661 The AIA-Pack BHCG Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK BHCG assay

DateSep 14, 2020 (initiated Aug 7, 2020)
StatusTerminated
ReasonDecreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.
Root causeUnder Investigation by firm
ActionThe firm, Tosoh Bioscience, issued an "URGENT: MEDICAL DEVICE RECALL" Letter dated August 7, 2020 via Fedex to its customers. The letter states reason for recall, health risk and action to taken: Check the lot numbers of the AIA - PACK HCG Calibrator Set and AIA-PACK HCG Calibration Verification Test Set against the lot numbers listed above. Lots not listed above are not within the scope of this recall and maybe used per AAM. Remove all affected AIA - PACK HCG Calibrator Sets and AIA - PACK HCG Calibration Verification Test Sets from your inventory. The products must be tallied including pr
Quantity39 boxes
DistributionWorldwide Distribution: US (Nationwide) and countries of: Bogota, Cayman Islands, Mexico, Peru, and San Salvador.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.