ClearanceBrief
Recall Z-0008-2021 ยท GDW Staple, Implantable

Covidien, LP

Endo GIA Auto Suture Universal Articulating Loading Unit 60 mm - 3.5 mm, Item Code 030458 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis.

DateAug 12, 2020
StatusOpen, Classified
ReasonThe device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Root causeUnder Investigation by firm
ActionUrgent Medical Device Recall notification letters dated 8/12/20 were sent to customers. Required Actions: 1.Please immediately quarantine and discontinue use of affected item codes with associated lot numbers listedon Attachment A. 2.Please return affected product as indicated below. All unused products from the affected item codes and associated lot numbers must be returned. 3.If you have distributed the Covidien Endo GIA" Auto Suture" universal articulating loading units listed on Attachment A, please promptly forward the information from this letter to those recipients. 4.Complete the Reca
Quantity900,007 total
DistributionWorldwide distribution - US Nationwide and Global Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111825ENDO GIA(TM) STAPLERS, DST SERIES(TM) GIA(TM) STAPLERS, DST SERIES(TM) TA(TM) STAPLERS
Covidien, Formerly US Surgical A Divison of Tyco H, Oct 25, 2011