Recall Z-2951-2020 ยท LLZ System, Image Processing, Radiological

SICAT GMBH & CO. KG

SICAT IMPLANT V2.0

DateJul 15, 2020
StatusTerminated
ReasonA dentist found implant positions are not correctly exported from the implant planning software SICAT IMPLANT V2.0 - for the specific export format CMG.DXD.
Root causeSoftware design
ActionOn 07/15/2020 customer notification letter was sent to all users/customers who have the 2.0 version of the software installed in the system. These users are instructed to only use the system for specific cases as defined in the field safety notice. The firm plans on implementing a new software version in July 2020 to correct the issues found.
Quantity853 units
DistributionWorldwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192348SICAT Implant V2.0
Sicat GmbH & Co. KG, Dec 4, 2019