Recall Z-2944-2020 ยท LLZ System, Image Processing, Radiological

Merge Healthcare, Inc.

Merge PACS

DateAug 6, 2020
StatusTerminated
ReasonMeasurements done on the Merge PACS generated MPR s may have incorrect measurements.
Root causeSoftware design
ActionThe firm, IBM Watson Health, sent an "Urgent Field Safety Notice" dated 05 August 2020 to its customers. The letter described the product, problem, and actions to be taken. The customers were instructed to do the following: Avoid incorrect pixel measurements by using the following workflow: Step 1: Modality acquires a series and generates reformat and sends to Merge PACS Step 2: Radiologist performs measurement on these series from the modality. Avoid black dots that may lead to additional imaging by taking any of the following actions: - Verifying study images either at modal
Quantity125 total units
DistributionDomestic Distribution: IL, MO, NY, ND, MI, TX, FL, OH, UT, NE, MS, MN, DE, AR, CA, WI, MD, NC, NH, GA, VA, SC, PE, CT, NM, IN. International Distribution: CA, BE, UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192455Merge PACS
Merge Healthcare Incorporated, Oct 17, 2019