Recall Z-2872-2020 ยท RCC C-Arm Fluoroscopic X-Ray System

Ge Healthcare

The OEC 9900 C-Arm is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical-care and emergency room procedures. The GE OEC 9800 is designed to provide fluoroscopic and spot-film imaging of t

DateAug 27, 2020 (initiated May 8, 2020)
StatusOpen, Classified
ReasonGEHC Surgery became aware that some users of the 9800 and 9900 fluoroscopic mobile C-Arms (referred to as the system here on) within the United States had requested that a Field Service Engineer (FSE) change the dose value on the live update screen from AKR / CAK to DAP. All systems sold to customers in the US that leave GEHC Surgery (the place of manufacture) are configured to display AKR / CAK when they leave the factory. An FSE changing the dose display, per the customer request, to display DAP instead does not comply to 21CFR 1020.32(k), which requires fluoroscopic equipment manufactured o
Root causeRadiation Control for Health and Safety Act
Quantity221
DistributionNationwide
FDA recordOpen on fda.gov