Recall Z-2921-2020 ยท DQK Computer, Diagnostic, Programmable

Siemens Medical Solutions USA, Inc

Siemens Sensis Vibe Hemo system in combination with the MicroPodTM EtCO2 module-diagnostic and administrative tool supporting hemodynamic catheterization and/or electrophysiology studies, for cardiac as well as interventional Radiology.

DateAug 28, 2020 (initiated Jul 24, 2020)
StatusTerminated
ReasonThe connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment conditions, the HemoBox may stop functioning and all vital signs are lost
Root causeProcess design
ActionSiemens issued Customer Safety Advisory Notice distributed dated July 24 2020 to customers via AX048/20/S. Letter states reason for recall, health risk and action to take: If the connector linking the EtCO2 module to the HemoBox should become damaged or break, the operator shall not attempt to fix the connector or re-connect the cable to the HemoBox. Only Siemens Customer Service representative or a service provider authorized by Siemens Healthineers may resolve the issue. Siemens strongly recommends that you do not attempt to fix the connector or re-connect the cable to the HemoBox until the
Quantity110 units US
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150493Sensis
Siemens Medi Cal Solutions, Inc., Jun 30, 2015