Recall Z-3016-2020 ยท LRP Tray, Surgical

Cardinal Health 200, LLC

KIT, PRE OP DRAH - Product Usage: Preoperative kit used to decolonize intranasal S. aureus.

DateJul 30, 2020
StatusTerminated
ReasonOnly providing one nasal swab in our pre-operative kit instead of the 4 required for proper treatment.
Root causeEmployee error
ActionOn July 30th the firm send its consignee a letter notifying them of the field action. Component Part Number: X12048 (SWABSTICK, PVP-I,10%, NASAL DECOLONIZATION, PROFEND,NS) is supplied to Cardinal Health, Presource in a package of 4 nasal swabs. Per the suppliers Instructions for Use, X12048 is a kit that needs to be administered in a 4-swab treatment with two swabs used per nasal passage. Presource is only providing one nasal swab in our pre-operative kit instead of the 4 required for proper treatment. The Presource kit was set up requiring 1 each of the swab component, but confusion in the c
Quantity9,150 kits
DistributionUS Nationwide distribution including the state of North Carolina.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.