Recall Z-2986-2020 ยท LWR Heart-Valve, Non-Allograft Tissue

Sorin Group Italia S.r.l.

Perceval Single Use Accessory Kit, Size S, REF: ICV1349, Sterile EO, Rx Only, CE, UDI: 8022057015211 - Product Usage: he Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation.

DateJul 29, 2020
StatusTerminated
ReasonLivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
Root causeNonconforming Material/Component
ActionLetter titled, Urgent Medial Device Removal, Perceval single-use accessory kit models assembly error, HV-SAL-2020-1, dated July 29, 2020 was sent to consignees via 1st class mail (certified return receipt) on July 29th, 2020 Informing the affected customers: o It could be impossible to collapse Perceval Sutureless heart valves (Perceval S and Perceval PLUS) while using specific lots of Perceval single-use accessory kits. The issue potentially affects kits assembled with a specific supplier lot; Actions to be taken by the Physician LivaNova is coordinating the removal and replacement of all
Quantity102 kits
DistributionUS Nationwide distribution including in the states of CA, FL, ID, ME, MN, MO, MS, NC, SD, OH, OR, PA, SD, TN, TX, VA, WA, WI and the countries of Canada, Japan, Germany, Netherland, Switzerland, Sweden, UK, Spain, Italy, France, Belgium, Finland, Greece, Slovenia, Malta, Poland, Romania, Croatia, Hu
FDA recordOpen on fda.gov