ClearanceBrief
Recall Z-2936-2020 ยท MYL Assay,Enzyme Linked Immunosorbent,Parvovirus B19 Igg

Diasorin Inc.

Parvovirus B19 IgG EIA kit Part No. V519IGUS

DateJul 17, 2020
StatusOpen, Classified
ReasonPrinted label inside the box lid of the Parvovirus has the incorrect Lot Specific Constant value
Root causeUnder Investigation by firm
ActionDiaSorin issued a customer notification letter on July 17, 2020. The letter states the following action to be taken: Please use the Lot Specific Constant value of 0.14 with your remaining inventory of Lot 71440040AA.. The Lot Specific Constant is used to calculate the cut-off value and use of the incorrect value may lead to: 1. An invalid test due to the Low Positive Control resulting above the upper limit of the assigned range. 2. The reporting of a false positive patient result if the Low Positive Control does not fail and invalidate the test run. If you have used the affected product
Quantity615 units
DistributionNJ, WI, NY, NC, MN, CA, IL, VA, MA, MI, CT Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.