Recall Z-2938-2020 ยท QLT Covid-19 Multi-Analyte Respiratory Panel Nucleic Acid Devices

Qiagen Sciences LLC

QIAstat-Dx Respiratory SARS-CoV-2 Panel REF 691223 (US IVD under FDA EUA) - Product Usage: intended for the qualitative detection and differentiation of nucleic acid from multiple respiratory viral and bacterial organisms, including the SARS-CoV-2 virus, in nasopharyngeal swabs (NPS) eluted in universal transport media collected from patients suspected of COVID-19 by their healthcare provider.

DateJul 15, 2020
StatusTerminated
ReasonThere is an increased rate of potential false positive results for the Influenza A (no subtype) target, which may lead to 1) inappropriate antiviral use, 2) erroneous or delayed differential diagnosis and missing of an alternative diagnosis.
Root causeMaterial/Component Contamination
ActionOn July 15, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers to inform them of the increased rate of potential false positive results for the Influenza A (no subtype) target. Customers were instructed to do the following: l -Please review and reassess all Influenza A (no subtype) positive samples tested with the affected lots. -Please note that as per Instructions for Use if only an Influenza A signal is present and no additional signal for any of the subtypes is generated, it can be due to either low concentration or, in very rare cases, a new varian
Quantity1474 kits
DistributionUS Nationwide distribution including in the states of AK, AL, CA, DC, DE, FL, GA, IL, IN, KY, MA, MD, MN, NC, NJ, NY, OH, OR, PA, TX, WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.