Recall Z-0259-2021 ยท FPA Set, Administration, Intravascular
ICU Medical, Inc.
5.5" (14 cm) Appx 0.39 ml, Smallbore Trifuse Ext Set w/3 MicroClave Clear, NanoClave T-Connector, 3 Clamps, Rotating Luer.1 unit per pouch, 50 multivac pouches per case. UDI: (01)10887709032153(17)241101(30)50(10)4475297; (01)10887709032153(17)250201(30)50(10)4711598. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration
| Date | Jul 6, 2020 |
| Status | Terminated |
| Reason | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Root cause | Process control |
| Action | On July 6, 2020, the firm, icumedical, sent an "URGENT: MEDICAL DEVICE RECALL" notification of the market action to direct and indirect consignees of record via UPS. Stericycle, Inc. was contracted to send letters to consignees. The following is included in the "Required Actions for Users" section of the medical device recall notification: 1) Discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the re |
| Quantity | Total of all products (Listed #1 thru 101) = 304735 units |
| Distribution | Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Sa |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K964435 | PRIMARY IV SET Icu Medical, Inc., Apr 2, 1997 |