Recall Z-2881-2020 ยท IYE Accelerator, Linear, Medical

Klarity Medical Products LLC

Klarity Vacuum Bags, Item Nos. R7504-30C, R7504-35BC,R7504-35C, R7644-32.5BC, RC7221-25BC, RC7504-17.5C - Product Usage: To be used by trained medical professionals for the stable support and positioning of patients undergoing external beam radiation therapy treatment in a clinic or hospital setting.

DateJul 31, 2020
StatusTerminated
ReasonVacuum bags potentially may not hold their vacuum adequately for needed patient treatment times.
Root causeComponent design/selection
ActionOn July 31, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers, informing them that vacuum bags may experience product failure wherein the bags fail to maintain an adequate vacuum over a required period of time. The loss of vacuum can lead to treatment inaccuracy due to shift of patient location. Customers were instructed to: 1. Inspect current inventory to locate any affected product. 2. Remove product from service. 3. Destroy and dispose of product (the firm recommended cutting up and disposing with normal trash). Klarity will replace affected bags or rei
Quantity111 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of CA, FL, MA, NY, PA and the country of UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132124KLARITY VACUUM BAG, KLARITY MOLD CUSHION
Klarity Medical Products, LLC, Nov 1, 2013