Recall Z-0229-2021 ยท FPA Set, Administration, Intravascular

ICU Medical, Inc.

78 IN (198cm) APPX 10.0 ml, 20 Drop Set 3-Port NanoClave Manifold, BCV, MicroClave. 1 unit per pouch, 24 pouches per case.UDI:(01)10887709090634(17)241201(30)25(10)4537328. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needl

DateJul 6, 2020
StatusTerminated
ReasonIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Root causeProcess control
ActionOn July 6, 2020, the firm, icumedical, sent an "URGENT: MEDICAL DEVICE RECALL" notification of the market action to direct and indirect consignees of record via UPS. Stericycle, Inc. was contracted to send letters to consignees. The following is included in the "Required Actions for Users" section of the medical device recall notification: 1) Discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the re
QuantityTotal of all products (Listed #1 thru 101) = 304735 units
DistributionWorldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Sa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K964435PRIMARY IV SET
Icu Medical, Inc., Apr 2, 1997