ClearanceBrief
Recall Z-0476-2021 ยท PAM Gram-Positive Bacteria And Their Resistance Markers

Luminex Corporation

Verigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP, IVD - Product Usage: IVD sample processor, which utilizes a single-use disposable test consumable and a self-contained Verigene Test Cartridge for each sample tested.

DateOct 13, 2020
StatusTerminated
ReasonLuminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENESP devices.
Root causeNonconforming Material/Component
ActionOn 10/13/2020, the firm sent the following notice: "As part of our corrective action plan, Luminex will be updating software on all VERIGENE SP instruments in the coming weeks, and systematically visiting every VERIGENE customer site to perform a supplemental preventative maintenance on your instruments. We will also likely be sending out a series of Customer Advisory Notifications with additional information on items related to our remediation efforts surrounding the Warning Letter, as well as updates related to other items identified for improved product performance. During this time, you ma
Quantity2838 units
DistributionWorldwide distribution. US nationwide and Kuwait, Austria, United Kingdom, Spain, France, Chile, Japan, Netherlands, Greece, Italy, Thailand, Germany, Turkey
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143653Verigene Respiratory Pathogens Flex Nucleic Acid Test (RP Flex)
Nanosphere, Inc., Sep 4, 2015