Recall Z-2910-2020 ยท FRN Pump, Infusion

CareFusion 303, Inc.

PC Unit Front Case with Keypad Replacement Kits: TC10008389 ASSY CASE FRONT W/KEYPAD 8015LS, TC10010217 ASSY FRT CASE W/ KEYPAD 8015 M2, TC10012515 ASSY FR CASE W/ KEYPAD 8015 M2, TC10013702 ASSY, CASE, FRONT W/KEYPAD, 8015LS, TC10013664 ASSY FR CASE W/ KEYPAD 8015 M2

DateSep 4, 2020 (initiated Aug 4, 2020)
StatusOpen, Classified
ReasonThe pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.
Root causeReprocessing Controls
ActionOn 08/04/20, Urgent Medical Device Recall notices were mailed to Clinical Users: If the PC unit keypad becomes unresponsive or a key gets stuck, clinicians should remove the device from service and send to Biomedical Engineering. If administering a critical medication, continue the infusion until it is safe to replace the PC unit. Cleaning personnel can minimize the potential for fluid ingress during cleaning by 1) not using a cloth that drips, 2) wringing out the cleaning cloth to squeeze out excess fluid, and 3) not spraying fluids directly onto the device. Biomedical Engineering: If an
Quantity28346
DistributionU.S.:LA, FL, KY, MI, AL, GA, CA, TN, NC, TX, MA, OH, CO, WI, IL, KS, MO, UT, AK, MD, NY, NV, VA, NE, IN, PA, VT, DC, SC, CT, AR, NJ, MT, ND ,HI , NH, WA, DE, ME, OR, MN, AZ, MS, IA, NM, WY, ID, OK, SD, WV, MP, GU O.U.S.: CA, AE, AU, BE, CA, CH, IL, IN, KW, N/A, NL, NZ, PH, QA, SA, SG, TW, US, ZA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133532ALARIS SYSTEM WITH GUARDRAILS SUITE MX
Carefusion 303, Inc., Aug 21, 2014